2
♦ The electrocardiograph is provided for the use of qualified physicians or personnel
professionally trained. And they should be familiar with the contents of this user
manual before operation.
♦ Only qualified service engineers can install this equipment. And only service engineers
authorized by our company can open the shell.
♦ Only qualified installation or service engineers can shift the mains shift switch
(100V~115V/220V~240V) according to local mains supply.
♦ The results given by the equipment should be examined with respect to the overall
clinical condition of the patient. And it can not substitute for regular checking.
WARNING :
♦ EXPLOSION HAZARD-Do not use the electrocardiograph in the presence of
flammable anesthetic mixture with oxygen or other flammable agents.
♦ SHOCK HAZARD-The power receptacle must be a hospital grade grounded outlet.
Never try to adapt the three-prong plug to fit a two-slot outlet.
♦ If the integrity of external protective conductor in installation or arrangement is in
doubt, the equipment should be operated from the built-in rechargeable battery.
♦ Do not use this equipment in the presence of high static electricity or high voltage
equipment which may generate sparks.
♦ This equipment is not designed for internal use and direct cardiac application.
WARNING :
♦ Only patient cable and other accessories supplied by our company can be used. Or else,
the performance and electric shock protection can not be guaranteed.
♦ Be sure that all electrodes have been connected to the patient correctly before
operation.
♦ Be sure that the conductive parts of electrodes and associated connectors, including
neutral electrode, should not contact with earth or any other conducting objects.
♦ Electrodes with defibrillator protection should be used while defibrillating.
♦ There is no danger for patients with pacemaker.
♦ Do not touch the patient, bed, table and the equipment while using defibrillator or
pacemaker simultaneously.