In order to avoid burning, please keep the electrode far away from the radio knife
while using electrosurgical equipment simultaneously.
WARNING :
Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective IEC standards (e.g. IEC 60950 for data processing
equipment and IEC 60601-1 for medical equipment). Furthermore all configurations
shall comply with the valid version of the standard IEC 60601-1-1. Therefore anybody,
who connects additional equipment to the signal input connector or output connector to
configure a medical system, must make sure that it complies with the requirements of
the valid version of the system standard IEC 60601-1-1. If in doubt, consult our
technical service department or your local distributor.
The summation of leakage current should never exceed leakage current limits while
several other units are used at the same time.
The potential equalization conductor can be connected to that of other equipment when
necessary, to make sure that all these equipment are connected with the potential
equalization bus bar of the electrical installation.
1.2.2 Battery Care Warnings
WARNING :
Improper operation may cause the battery to be hot, ignited or exploded, and it may
lead to the declination of battery’s capacity. It is necessary to read the user manual
carefully and pay more attention to warning messages.
Only qualified service engineer authorized by our company can open the battery
compartment and replace the battery. And the battery of same model and specification
provided by manufacturer should be used.
Danger of explosion -- Do not reverse the anode and cathode when connecting the
battery.
Do not heat or splash the battery or throw it into fire or water.
When leakage or foul smell found, stop using the battery immediately. If your skin or
cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If
the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean
water first and go to see a doctor immediately.