3.3. Clinical restrictions
A. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of the
testee is required. For a testee with weak pulse due to shock, low ambient/body temperature,
major bleeding, or use of vascular contracting drug, the SpO
2
waveform (PLETH) will decrease.
In this case, the measurement will be more sensitive to interference.
B. For those with a substantial amount of staining dilution drug (such as methylene blue, indigo
green and acid indigo blue), or carbon monoxide hemoglobin (COHb), or methionine (Me+Hb)
or thiosalicylic hemoglobin, and some with icterus problem, the SpO
2
determination by this
monitor may be inaccurate.
C. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major
factor blamed for serious error of SpO
2
measure.
D. The SpO
2
value serves only as a reference value for judgement of anemic anoxia and toxic
anoxia, some patients with serious anemia may also report good SpO
2
measurement.
4. Technical specifications
A. Display mode: 65K Color OLED display
SpO
2
measuring range: 35%~100%
Pulse rate measuring range: 30bpm~240bpm
Pulse waveform display: bar graph display
Battery indication: OLED Signal
B. Power supply requirement:
Lithium battery × 1
Adaptable range: 3.5V~4.2V
C. Operating current: ≤80mA
D. Resolution: SpO
2
: 1%
Pulse Rate : 1bpm
E. Accuracy: SpO
2
:70% to 100%±2 digits, below 70% unspecified
Pulse rate:±2 bpm or ±2% (select larger)
F. Measurement at low perfusion:
The values of SpO
2
and pulse rate can be displayed properly when pulse saturation is at 0.4%.
Accuracy of SpO
2
:±4%
Accuracy of Pulse rate:±2bpm or ±2% (select larger)
G. Resistance to surrounding light:
The difference between the value measured in the condition of man-made light and indoor natural