Cryo-con Model 24C Remote Programming Guide
SYSTem:RESeed
Re-seeds the input channel’s averaging filter, allowing the reading to settle
significantly faster. The display filter may have filter time-constants that are
very long. The RESEED command inserts the current instantaneous
temperature value into the filter history, thereby allowing it to settle rapidly.
i Note: The RESEED command is useful in systems where a
computer is waiting for a reading to settle. Issuing the RESEED
command will reduce the required settling time of the reading.
SYSTem:HOMe
Causes the front panel display to go to the Operate Screen.
SYSTem:SYNCtaps <taps>
Sets or queries the number of taps in the synchronous filter. This is an
advanced setup function. The default is 7 taps.
SYSTEM:NAME "name"
The controller contains a unit name string that may be set or queried using
this command. This can be used to assign a descriptive name to the
instrument.
SYSTem:HWRev?
Queries the instrument’s hardware revision level.
SYSTem:FWREV?
Queries the instrument’s firmware revision level.
SYSTem:LINefreq {60 | 50}
Sets or queries the AC Power Line frequency setting which may be either 50
or 60 for 50Hz or 60Hz. Command only affects the operation of the
synchronous cryo-cooler filter.