Cryo-con Model 24C Remote Programming Guide
INPut {A | B | C | D}:ALARm:LOENa { YES | NO }
Sets or queries the low temperature alarm enable for the specified input
channel. An alarm must be enabled before it can be asserted.
INPut {A | B | C | D}:ALARm:LTENa { YES | NO }
Sets or queries the latched alarm enable mode. When an alarm is latched, it
can be cleared by using the CLEar command.
INPut {A | B | C | D}:ALARm:CLEar
Clears any latched alarm on the selected input channel.
INPut {A | B | C | D}:ALARm:AUDio { YES | NO }
Sets or queries the audio alarm enable. When enabled, an audio alarm will
sound whenever an alarm condition is asserted.
INPut {A | B | C | D}:MINimum?
Queries the minimum temperature that has occurred on an input channel
since the STATS:RESET command was issued.
INPut {A | B | C | D}:MAXimum?
Queries the maximum temperature that has occurred on an input channel
since the STATS:RESET command was issued.
INPut {A | B | C | D}:VARiance?
Queries the temperature variance that has occurred on an input channel
since the STATS:RESET command was issued. Variance is calculated as the
Standard Deviation squared.
INPut {A | B | C | D}:SLOpe?
Queries the input channel statistics. SLOPE is the slope of the best fit straight
line passing through all temperature samples that have been collected since
the STATS:RESET command was issued. SLOPE is in units of the input
channel display per Minute.
INPut {A | B | C | D}:OFFSet?
Queries the input channel statistics. OFFSET is the offset of the best fit
straight line passing through all temperature samples that have been
collected since the STATS:RESET command was issued. OFFSET is in units
of the input channel display.
INPut {A | B | C | D}:STAts:TIMe?
Queries the time duration over which input channel statistics have been
accumulated. Time is reset by issuing the STAt:RESet command. Query only.
INPut {A | B | C | D}:STAts:RESet
Resets the accumulation of input channel statistical data. Command only
affects the selected input channel.
INPut {A | B | C | D}:TCOFfset <offset>
Sets or queries the offset value for thermocouple inputs. <offset> is the
decimal value of offset and is in units of Kelvin. Refer to the section on Using
Thermocouple Sensors for more information.