Cryo-con Model 24C Remote Programming Guide
Data Logging Commands
DLOG:STATe {ON|OFF}
Turns the data logging function ON or OFF. Equivalent to Start / STOP.
DLOG:INTerval <Seconds>
Sets the data logging time interval in seconds.
DLOG:COUNt?
Queries the number of entries in the log buffer.
DLOG?
DLOG:READ?
Reads the entire contents of the log buffer. Each record is sent on a single
line. Format is:
<#>, MM/DD/YYYY, HR,MN,SC, ChA, ChB, ChC,ChD
where:
<#> is the record number.
MM/DD/YYYY is the date in Month, Day, Year format.
HR,MN,SC is the time in Hour, Minute, Second format.
Lines end with a <CR><LF> sequence. End of transmission is indicated by a
line that only contains a semi-colon.
DLOG:RESEt
Sets the logging record number to zero.
DLOG:CLEAr
Clears the data logging buffer.