Cryo-con Model 24C Specifications, Features and Functions
Relay Outputs
Number: 2
Type: Dry-contact.
Contact Rating: 10A @125 VAC, 5A @250 VAC or 5A @30 VDC for
resistive loads.
Function: Asserted or cleared based on temperature setpoint data.
Deadband: User defined.
Connector: 10-pin detachable terminal block.
Remote Interfaces
Remote interfaces are electrically isolated to prevent ground loops.
Ethernet: Supported protocols include: HTTP, TCP/IP, UDP and SMTP.
Electrically isolated.
RS-232: Serial port is an RS-232 standard null modem. Rates are 9600,
19,200, 38,400, 557,600 and 115,200 Baud.
IEEE-488 (GPIB): External option. Full IEEE-488.2 compliant.
USB 2.0: External option. Serial port emulator.
Language: Remote interface language is IEEE-488.2 SCPI compliant.
Further, it is identical within the entire Cryo-con instrument line.
Compatibility:
National Instruments LabView™ drivers available for all interfaces.
Ethernet API available for C++ and Basic.
User Setups
Four User Setups are available that save and restore the complete
configuration of the instrument.
General
Ambient Temperature: 25
o
C 5
o
C for specified accuracy.
Mechanical: 8.5"W x 3.5"H x 12"D. One half-width 2U rack. Instrument bail
standard, rack mount kit optional.
Weight: 9 Lbs.
Enclosure: Aluminum. Machined Aluminum front panel.
Power Requirement: 100, 120, 220 or 240VAC +5% -10%.
50 or 60Hz, 150VA max.
26