Cryo-con Model 24C Basic Setup and Operation
To set the Synchronous Filter Taps parameter, enter a number between 1 and 25 into
the Sync Filt. Taps field. A setting of 1 turns the filter off.
Most cryocoolers use a setting of 7 since this is the most common sub-multiple of the
AC line frequency applied.
i Note: If you are not using a cryocooler, please leave the Sync
Filt. Taps field set at the default of 7.
i Note: If you change the setting the Sync Filt. Taps setting, you
will need to re-tune the PID control loop.