GENERAL DESCRIPTION
Page 2-14
* Tolerances are based upon the full temperature range of the chamber. Better control will be achieved across a limited range.
(SOUND MEASURED 3FT (.91 m )FROM UNIT & 3FT (.91 m) ABOVE FLOOR IN dba)
(Contact factory for sound data on the other models not listed here.)
Table 1. ZP(H) Specifi cations
ZP(H) - 8 ZP(H) - 16 ZP(H) - 32 ZP(H) - 44 ZP(H) - 64 ZP(H) - 80
Workspace
Volume
8 Cubic Ft (230 L) 16 Cubic Ft (450 L) 32 Cubic Ft (900 L) 44 Cubic Ft (1,250 L) 64 Cubic Ft (1,810L) 80 Cubic Ft (2,265L)
Exterior
Dimensions
36”W x 57”D x 76”H
(91cm x 145cm x 193cm)
42”W x 63”D x 82”H
(107cm x 160cm x 208cm)
50”W x 71”D x 91.5”H
(127cm x 180cm x 232cm)
56.5”W x 71”D x 99.5”H
(143.5cm x 180cm x 253cm)
60.5”W x 81”D x 101.5”H
(154cm x 205cm x 258cm)
67”W x 94”D x 101.5”H
(170cm x 238cm x 258cm)
Workspace
Dimensions
24”W x 24”D x 24”H
(61cm x 61cm x 61cm)
30”W x 30”D x 30”H
(76cm x 76cm x 76cm)
38”W x 38”D x 38”H
(97cm x 97cm x 97cm)
44”W x 38”D x 46”H
(112cm x 97cm x 117cm)
48”W x 48”D x 48”H
(122cm x 122cm x 122cm)
48”W x 60”D x 48”H
(122cm x 152cm x 122cm)
Temperature
Ranges
*Temperature
Control
Tolerance
±0.5°C at steady state condition after stabilization
Humidity Range
Optional Range
*Humidity
Control
Tolerance
±3% RH at steady state conditions after stabilization
Distributed
Shelf Load
Capacity
110 lbs. 110 lbs. 100 lbs. 100 lbs. 100 lbs. 100 lbs.
Live Load Capacity Electrical Power Requirements
Watts Full Load Amps ¹
Model -18°C -34°C 208-230V, 1Phase 208-230V, 3Phase
ZP-8-2-H/AC 1200 250 34 30
ZP-16-2-H/AC 1200 250 34 30
ZP-32-2-H/AC 1200 250 34 30
ZP-44-2-H/AC 2000 300 43 38
ZP-64-2-H/AC 2000 300 43 38
Table 2-4. ZP Load Capacity and Electrical Requirements
Electrical PowerRequirements
Single Stage -34°C to 190°C