Status / messages
Ergometer ec
3000
10
-
2
10.2.
List of messages
10.2.1.
Messages of the blood pressure measuring system
In the following messages of the blood pressure measuring system are listed. The
message is described and notes for the elimination of the cause are given.
Error number
D
escription
Notes for the elimination
900
Common error message
Different causes
( visual inspection )
904
The limit value absolutely
exceeds the blood pressure
measuring values
Heart rate < 35 1/min
Systole < 70 mmHg
Diastole < 40 mmHg
Systole
–
Diastole
< 15 mmHg
905
Relief rate outside of tolerance
Cuff leaky ( defect )
Valves leaky
906
Faulty measuring
Too much exercise artefacts
910
Zero point deviation too much
System not without pressure
when it is switched on
916
Pressure rise too fast
Cuff tub
e dejected,
cuff defect
917
Pressure rise too slow
Cuff leaky ( defect )
Valves leaky
918
No diastole pulses
up to 150 mmHg
Cuff wrongly applied or
oscillations too weak
920
Not enough oscillations
recognized
Cuff wrongly applied or
oscillations too we
ak
929
930
Not enough oscillations after
artefact extraction
Cuff wrongly applied or
oscillations too weak
936
Sensor error pressure
measuring
Service / repair required
937
max. cuff pressure exceeded
Repeat measuring