WP
4
PotentiaMeter
Menus
21
menu. For more details on calibration and how to
verify it, please refer to the next chapter.
Setting the Temperature (WP4-T only)
If you have a WP4-T, you have the ability to manu-
ally set your instrument’s sample chamber tempera-
ture. This is done by pressing the lower right button
next to the “
TE set
” in the system configuration
menu. The following screen will appear:
Adjusting the Setpoint Temperature
Use the buttons next to the + and - buttons to
adjust the target setpoint temperature (displayed in
the lower right corner). If you press either button it
adjusts in increments of .5°C. Note: Holding down
the button will not increment the value, so you must
press the button each time you wish to change this
value.
The target setpoint temperature roughly corre-
sponds to the temperature at which you wish the
chamber to read. Therefore, adjust the setpoint to
the temperature that you want, then begin measure-
ments to see how close your WP4-T comes to your
desired temperature (this works best by putting the
WP4 in continuous mode). After several samples, it
0 Adjust setpoint
-Exit-
25.0
+
-