WP
4
PotentiaMeter
Troubleshooting
57
1. PROBLEM:
WP4 won’t turn on.
SOLUTION:
2. Long read time Dirty sample chamber
Sample desorbs slowly
Broken chamber fan
blade
3. Readings on KCl
standards are too high/
low to adjust
Dirty thermopile
If using own salt solu-
tion, it may not be in
equilibrium
4. “Sample too dry” Sample too dry to accu-
rately measure
Dirty chamber or mirror
5. Triangle appears in
upper right corner
Mirror is dirty
6. “Block ROM failed”
appears on screen after
turning on WP4
Block is disconnected
from motherboard
Block memory compo-
nent has failed
7. “TE set” option no
longer appears on
screen
Temperature control mod-
ule in WP4 is broken or
not functioning
Table 1: Troubleshooting
Problem Possible Solutions