WP
4
PotentiaMeter
Troubleshooting
59
Decagon to make arrangements for repairs.
2. PROBLEM:
Readings are slow or inconsistent.
SOLUTION:
The sample chamber may be dirty. Refer to
Chapter 10 of the manual for directions on
cleaning the sample chamber.
Some samples absorb or desorb moisture
very slowly, causing measurements to take
longer than usual, and nothing can be done to
speed up the process. Refer to Chapter 6 for fur-
ther explanation.
A fan blade may be broken inside the block.
If even the KCl standard takes a long time to
read, and the sample chamber is clean, you may
have a broken chamber fan blade. This is espe-
cially likely if you have just cleaned the cham-
ber. If you suspect this may have happened,
contact Decagon for details on replacement.
3. PROBLEM:
Water potential readings on KCl standards are
too high/low and a calibration adjustment cannot
be made any higher/lower.