11
3 TECHNICAL PARAMETERS
3.1 Electrical impulses of the device have the following output parameters:
3.1.1 When load = 20 kOhm
Capacity (in standard units)
Minimum* Maximum**
1 99
Amplitude of phase 1, B ≤ 40 ≤ 40
Duration of phase 1, ms 6,0 ± 1,5 400 ± 100
Amplitude of phase 2, B 9,0 ± 2,5 450 ± 113
The above stated parameters apply to treatment mode impulses and do not apply to test impulses.
Parameters should be measured when Therapy without Contact Control Function is enabled.
Parameters should be measured when Children’s Doctor Function is disabled.
*
**