22 3.12 Recommendations for determining the necessary distance between DENAS-PCM
Electrical Stimulator and radio equipment.
Label maximum output capacity
of the transmitter P (W)
Radiation frequency and formula to determine the distance
d (m)
150 kHz – 80 МHz
d = 1,2√P
150 kHz – 800 МHz
d = 1,2√P
800 МHz – 2,5 GHz
d = 2,3√P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
3.13 Conditions of Use.
3.13.1 Connection of external electrodes not-authorized by the manufacturer
(including other cable length, cable type, electrode construction) may lead to increase