Denver Instrument MAXX - Balance Leveling and Calibration Preparation; Leveling, Weight Access, and Warm-up

Denver Instrument MAXX
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
Leveling the Balance/Scale
(only for model MXX-123)
§ Always level the balance/scale
again any time after it has been
moved to a different location.
§ Turn the front feet as shown in the
diagram until the air bubble is cen-
tered within the circle of the level
indicator.
§ Lower all feet to touch the counter-
top so the balance/scale does not
rock.
> In most cases this will require
several adjustment steps.
Removing Weights for
Calibration/Span Adjustment
(only for models MXX-612, MXX-412,
MXX-212, MXX-123)
§ Grasp the tab to turn and pull out
the weight compartment.
§ Always wear gloves or use tweez-
ers when handling weights.
$ Follow instructions on page 15 for
calibration/span adjustment.
Warmup Time for Model MXX-123:
To ensure accurate results, the scale
must warm up for 1 hour before
operation. Only after this time will
the scale have reached the required
operating temperature.