neoprobe
TM
GDS Operation Manual
AW-000639 Page 16 of 38
Ratio Readout [9] is displayed. Ratio Symbol [12] flashes.
After 5 seconds, system returns to Binary Pitch Mode displaying last calculated Background Count data [8]
and current Count Rate data [7].
TARGET COUNT; 6-SECOND OPTION
Operation Note
Events that may occur during product use:
If…..
Then this occurs..
Target Count Button [11] is activated
(press and hold) while a Target Count is
calculated
Start Beep sounds and starts new Target
Count
Target Count [7] is too high
Overflow symbol ///// is displayed in
Target Count Readout (7) and beep
sounds
/////
Target Count button [11] is pressed twice
while Target Count Readout data is
displayed
Starts new Target Count
Calculated Ratio Readout data is greater
than 99.9
Overflow symbol ///”appears in Ratio
Readout (9)
///
Performing a Target Check (2-Second Option)
Perform a 2-second Target Check to obtain a quick reading of radioactivity level of target tissue.
System status before Target Check
Background Count Readout [8] displays Background Count Data.
Ratio Readout [9] is off.
Target Count [7] displays count rate as probe is moved over tissue.
How to perform Target Check
Hold probe stationary over target tissue.
Press and release Target Count Button [11]; Start Beep sounds.
Hold probe steady for 2 seconds until Stop Beep sounds.
Console Display
Target Count Symbol [10] flashes.
Target Count Readout [7] displays dashes.
Artwork No: AW-000639 Revision: A Artwork Status: Released Lifecycle Name: Artwork
PCO No: N/A ECN No: ECN-000214 Release Date: 10/09/2012 Effective Date: 10/09/2012