neoprobe
TM
GDS Operation Manual
AW-000639 Page 21 of 38
External collimator is removed from probe.
Approved Cleansers/Sterilizers
The console is compatible with following cleaning and disinfection agents/processes:
Control III (Maril Products, Inc.) – probes cables, collimators & connectors
Metrizyme (Metrex Research Corporation) – probes cables, collimators & connectors
Mild Soap and Water Solution – all products
Tor-II (Huntington Laboratories) – all products
Ammonia and 97% water – all products
Clorox and 90 to 95% water – all products
MediClean Forte – all products
MediZym – all products
Isopropyl Alcohol 70% (rubbing alcohol) – all products
The accessories are compatible with the cleansers listed above and the sterilizers listed below. Refer to individual
Cleaning and Sterilization Instruction for Use Documents on Manual CD included with product for detailed
instructions. See list of documents below:
Cleaning/Sterilization Method
Cleaning and Disinfection of probes, cables and collimators according to EN ISO 17664:2004
STERRAD
®
SYSTEM Reprocessing of probes, cables and collimators according to EN ISO
17664:2004
Ethylene Oxide (EO) Reprocessing of probes, cables and collimators according to EN ISO
17664:2004
Autoclave Reprocessing of collimators according to EN ISO 17664:2004
NOTE: cleaning agents not listed above are sole responsibility of user to adequately assess and validate.
CAUTION: When using radioactive materials, use safe and proper handling techniques. See your
institution’s radiation safety officer about nuclear regulation commission and other requirements.
Radioactive decontamination procedures
An increase in counts without a known radioactive source may indicate radioactive contamination of environment,
probe or accessories. If high count is detected in absence of a radioactive source, then probe and collimator should be
cleaned with a radioactive decontaminant solution (e.g., Radiacwash®).
Follow standard nuclear medicine decontamination techniques. Wash probe with a decontaminant solution (e.g.
Radiacwash®) using a soft gauze pad. Pay particular attention to cleaning recesses, crevices and mating surfaces
with a soft brush. Do not scratch or abrade surfaces. Treat spent cleaning solution as radioactive waste and do not
allow it to contaminate other surfaces. Dispose of pads and cleaning solution in approved containers.