neoprobe
TM
GDS Operation Manual
AW-000639 Page 28 of 38
Probe with BLUETOOTH
®
wireless
technology: Unit displays “no linc” on
power up
Replace probe battery and
reactivate
Battery is too low to operate probe
Replace battery
Return for service if condition persists
For cabled probe: Unit displays “error no
probe” on power up.
Probe or cable not connected
Return for service if condition persists
Unit displays “Error 2” through
“error 6”
Cycle power to unit
Return for service if condition
persists
Intermittent and unusually high readings
Intermittent electrical interference
(e.g., electrosurgical unit, X-ray,
etc.)
Track down source of interference
Turn off offending sources (if
possible)
Change probe
Return for service if condition
persists
Unexpected high readings (continuous)
Contaminated probe or collimator
Decontaminate probe and
collimator
Incorrect Radionuclide selected
Select appropriate radionuclide
Return for service if condition
persists
Incorrect Radionuclide selected
Select appropriate Radionuclide
Change probe
Return for service if condition
persists
Erratic readings or no counts
Return for service if condition
persists
Sound does not vary with count rate
Wrong count range is active
Lower count range
Return for service if condition
persists
Artwork No: AW-000639 Revision: A Artwork Status: Released Lifecycle Name: Artwork
PCO No: N/A ECN No: ECN-000214 Release Date: 10/09/2012 Effective Date: 10/09/2012