SH_TSH not found: all readings have a light value too low. The sensor is probably
faulty
SH_TSH found too low: a light value above SH_TSH was found too low, probably
beyond the bottom of the tube. The tube is too full or obscured
Blood level not found with method M1
Blood level over SL_TSH: following the first reading, the blood level is too low
The blood level detected at the second reading is higher than that of the first
reading
The ESR result is equal to 0 (allowed values from 1 to 140)
The ESR result is higher than 140
Tube removal detected during reading
The control method (CM) failed to detect the blood level
The blood levels found by M1 and the Control Method (CM) differ more than the
acceptability value
Unexpected test tube: error related to the insertion of an anomalous tube
*The -10 is not an error but denotes that the value of ESR is above 140 and is displayed with ">
140"