Rev. 1.2 (12.2021) MINI-CUBE | SERVICE MANUAL 39
4
ERR
SH_TSH not found: all readings have a light value too low. The sensor is probably
faulty
5
HIGH
SH_TSH found too low: a light value above SH_TSH was found too low, probably
beyond the bottom of the tube. The tube is too full or obscured
6
ERR
Blood level not found with method M1
7
LOW
Blood level over SL_TSH: following the first reading, the blood level is too low
8
ERR
The blood level detected at the second reading is higher than that of the first
reading
9
ERR
The ESR result is equal to 0 (allowed values from 1 to 140)
10
*
>140
The ESR result is higher than 140
11
ERR
Tube removal detected during reading
13
ERR
The control method (CM) failed to detect the blood level
14
ERR
The blood levels found by M1 and the Control Method (CM) differ more than the
acceptability value
15
ERR
Unexpected test tube: error related to the insertion of an anomalous tube
*The -10 is not an error but denotes that the value of ESR is above 140 and is displayed with ">
140"