Operation
Wait
for
warming-up
time.
Wait
for
about
35
minutes
before
use.
Heat
incubator
in
"air
temperature
control"
operating
mode.
If
incubator
has
to
be
used
urgently:
Operate
incubator
in
standby
mode:
Set
air
temperature
to
32
°C
to
36
°C,
see
p.
20,
and
set
humidity
control
to
0,
see
p.
30.
Just
before
placing
the
baby
in
the
incubator:
Fill
water
container
If
required
use
protective
gloves.
1
Take
cover
of
water
container
off
and
2
fill
container
to
mark
with
clean,
distilled
or
demineralised
water
(hospital
dispensary).
Do
not
use
any
additives.
Take
care
not
to
overfill.
Put
cover
of
water
container
back.
Then
disinfect
hands.
If
water
bottles
are
used:
Fill
water
bottles
If
required
use
protective
gloves.
1
Open
back
panel
and
fold
down.
2
Disconnect
empty
water
bottles
and
refill
with
500
clean,
distilled
or
demineralised
water
(hospital
dispensary).
Do
not
use
any
additives.
Refit
filled
water
bottles
and
close
back
panel
again.
Then
disinfect
hands.
Placing
baby
in
the
incubator
Open
front
door
and
pull
out
bed.
Put
the
baby
on
the
bed
and
slide
the
bed
back
in
position.
Close
the
front
door
and
check
that
the
catches
are
properly
engaged.
Adjust
the
bed
as
required.
19