Dräger 8000 NC - Cleaning;Disinfecting;Sterilizing.35

Dräger 8000 NC
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Care
Stripping
down
Cleaning/disinfecting/sterilizing
1
Remove
mattress
from
bed.
2
Lift
off
bed.
Undo
screw
on
ventilator
hose
support
and
remove
hose
support.
Remove
bed
extension.
3
Pull
both
support
brackets
up
and
out.
4
Remove
base
plate
Caution.
Be
careful
not
to
burn
yourself
The
risk
of
burns
from
the
heater
is
particularly
great
when
the
incubator
has
been
closed
and
this
risk
remains
for
a
long
time
after
the
unit
has
been
switched
off.
(It
takes
about
one
hour
for
the
temperature
to
fall
to
70
°C).
5
Remove
fan.
6
Remove
trough.
Cleaning/disinfecting/sterilizing
For
users
in
the
Federal
Republic
of
Germany
we
recom¬
mend
that
only
disinfectants
on
the
current
DGHM
(DGHM:
German
Society
for
Hygiene
and
Microbiology)
list
of
surface
disinfectants
are
used.
For
medical
and
hygienic
reasons,
disinfectants
based
on
quaternary
compounds
should
be
used.
Damage
may
occur
to
the
material
of
the
incubator,
parti¬
cularly
to
the
sensetive
covering
parts
(PMMA/plexiglass
and
PC/polycarbonate),
if
preparations
are
used
which
are
based
on
-
chlorine-releasing
compounds
-
organic
and
inorganic
acids
-
alcohols.
With
regard
to
material
compatibility
we
recommend
the
following
surface
disinfectants:
Demykosan
AF
Messrs.
Bayrol,
Munich
Incidur
Messrs.
Henkel,
Dusseldorf
Sekusept
Powder
Messrs.
Henkel,
Dusseldorf
Follow
manufacturer's
instructions.
A
precondition
for
using
non-recommended
disinfectants
is
that
the
manufacturer
can
prove
material
compatibility
with
PMMA
and
PC.
35