Care
Cleaning/disinfecting/sterilizing
Canopy,
inside
and
outside
Inner
wail
folded
down
Mattress
Bed
/
Bed
extension
Support
brackets
Base
plate
Trough
Incubator
housing,
inside
and
outside:
Remove
obvious
soiling
with
disposable
cloth
and
detergent.
Wipe-disinfect
surfaces.
Allow
time
for
the
disinfection
process,
then
wipe
the
surfaces
again
with
a
clean,
damp
cloth
and
dry.
Fan
Seals
Hose
seals
Water
hose
Water
container/Bottle
holder:
Wash
with
detergent
and
rinse
with
clean
water.
Disinfect
in
bath.
Allow
time
for
the
disinfection
process,
then
rinse
with
clear
water
and
dry;
or
sterilize
at
120
°C
(glove
programme).
Use
recommended
cleaning
and
disinfecting
agents
only.
If
other
agents,
such
as
alcohol
for
instance,
are
used
there
is
a
risk
of
tension
cracks
in
the
Acrylic
and
Makrolon.
Do
not
use
UV-radiation
on
the
incubator.
This
may
cause
cracks
in
the
acrylic
parts.
36