Care
Before
re-using
Before
re-using
with
a
patient
Re-assemble
equipment,
see
"Stripping
down",
p.
35.
Fitting
seals
in
the
hand
port
openings:
Open
hand
port.
Insert
seal
marked
with
an
"L"
on
nose,
into
left
opening.
Insert
seal
marked
with
an
"R"
on
nose,
into
right
opening.
Position
nose
of
seal
on
hinge.
Put
sealing
lip
(thin
edge
of
profile)
to
the
outside.
Close
hand
port.
Testing
Readiness
for
Operation,
see
p.
14.
Allow
water
heater
to
run
dry:
-
to
disinfect
heater
-
to
test
that
"water
shortage"
display
is
working
properly.
Select
humidity
setting
»5«,
see
p.
30.
Allow
heater
to
run
until
"water
shortage"
alarm
is
triggered,
see
p.
30.
Switch
off
humidity
control
=
humidity
setting
»0«.
Do
not
refill
water
container
until
just
before
placing
baby
in
incubator.
If
a
bottle
holder
is
used:
Do
not
put
full
bottles
into
bottle
holder
until
just
before
placing
baby
in
incubator.
Operate
incubator
in
standby,
p.
19
or
switch
off
incubator,
cover
with
dust
cover
and
store
ready
for
use.
37