Maintenance
Intervals
Disposal
of
NiCd
batteries
Maintenance
Intervals
Clean
and
disinfect
incubator
and
parts
before
carrying
out
any
maintenance*
procedures,
and
before
returning
for
repair.
Air
filter
replace
after
two
months,
p.
15.
Dispose
of
as
normal
waste.
Seals
of
hand
port
openings
replace,
when
the
material
becomes
brittle
or
tacky
or
when
a
proper
seal
can
no
longer
be
obtained.
Hose
seals
replace,
when
the
material
becomes
brittle
or
tacky
and
when
the
lamination
is
torn.
Ventilator
motor
grease
with
10
drops
of
51524-HLP
32
oil
every
six
months
by
trained
service
personnel.
NiCd
battery
for
power
failure
replace
once
a
year
by
trained
service
personnel.
Disposal,
see
below
p.
42.
Inspection
and
service*
every
six
months
by
trained
service
personnel.
Disposal
of
NiCd
batteries
-
do
not
throw
into
fire;
NiCd
batteries
may
explode.
-
do
not
open
by
force;
risk
of
corrosion.
Batteries
must
be
treated
as
special
waste:
in
accordance
with
local
waste
disposal
regulations.
Information
can
be
obtained
from
local
environmental
and
public
health
authorities
or
from
approved
waste
disposal
companies.
According
to
DIN
31
051
the
following
definitions
apply:
Inspection
=
determining
actual
condition
Service
=
measures
to
maintain
required
condition
Repair
=
measures
to
re-establish
required
condition
Maintenance
=
inspection,
service
and,
when
necessary,
repair
42