Operation
Phototherapy
Interface
Using
phototherapy
unit
800/8000/4000
Accessory
Follow
relevant
Instructions
for
Use.
Push
trolley
carrying
phototherapy
unit
under
incu¬
bator
or
in
the
case
of
Photo-Therapy
4000
can
also
be
placed
directly
on
top
of
the
canopy.
Adjust
height:
Minimum
distance
between
bottom
of
phototherapy
unit
and
patient
is:
0.4
m
for
Photo-Therapy
800/8000
0.3
m
for
Photo-Therapy
4000
If
the
height
of
the
bed
is
changed
later,
make
sure
that
this
minimum
clearance
is
still
maintained.
Follow
safety
advice,
see
p.18.
Exampel:
Photo-Therapy
4000
Using
"BabyLink
Incubator"
interface
(optional)
The
"Babylink
Incubator"
interface
provides
incubator
data
(actual
values,
set
values,
alarms)
via
a
serial
interface.
At
present,
measured
values
can
be
displayed
as
trends
on
the
Drager
Babyguard
8000
Paediatric
Monitor.
Other
possible
connections
are
the
Drager
BabyScreen
or
any
Windows-PC
which
can
run
the
ThermoView
program.
The
interface
protocol
is
included
in
the
BabyLink
modification
set
(82
90
607).
Follow
Instructions
for
Use
for
interface.
33