Intended
Use
Intended
Use
Heat
therapy
for
premature
babies
and
sick
neonates;
with
control
of
air
temperature
and
humidity;
and
also
skin
temperature
as
an
option.
The
incubator
must
only
be
used
by
properly
trained
staff
under
the
supervision
of
qualified
medical
staff
who
have
up-to-date
knowledge
of
the
risks
and
benefits
of
the
use
of
incubators.
Do
not
use
the
skin
temperature
control
on
babies
who
are
in
shock
or
who
have
high
temperatures.
The
air
in
the
incubator
should
only
be
enriched
with
oxygen
when
prescribed
by
a
doctor.
It
is
absolutely
essential
that
such
oxygen
enrichment
is
controlled
on
the
basis
of
the
arterially-measured
oxygen
partial
pressure
in
the
patient's
blood.
If
this
is
not
done
there
is
a
risk
of
hyperoxaemia
(damage
to
the
eyes)
and
hypoxaemia
(damage
to
the
brain).
Do
not
nebulize
any
medicaments
or
similar
substances
in
the
patient's
room.
If
nebulized
substances
fall
onto
the
incubator
this
may
impair
its
functioning.
Mobile
telephones
must
not
be
used
within
10
metres
of
the
incubator.
Mobile
telephones
may
interfere
with
the
functioning
of
electro-medical
equipment.
Maximum
load
must
not
exceed
25
kg.
Only
connect
incubator
to
a
mains
power
socket.
Do
not
use
a
mains
distribution
board.
If
there
is
a
circuit
breaker
in
the
supply
to
the
distribution
board
the
permitted
limit
for
leakage
currents
might
be
exceeded.
There
could
then
be
an
electrical
risk
for
patient
and
staff.
4