Fault,
Cause,
Remedy
Main
module
Fault,
Cause,
Remedy
Main
module
Fault
Cause
Remedy
Red
Alarm
LED
is
lit.
Alarm
tone
commences.
Fault
in
module.
Identify
faulty
module.
See
"Fault,
Cause,
Remedy"
of
module.
Red
Inop.
LED
is
lit.
Continuous
tone
commences.
Basic
fault
in
electronics.
Switch
incubator
off
and
on
again.
If
Inop
message
not
repeated:
press
(X)
key
and
reset
values.
If
Inop
message
repeated,
incubator
not
working.
Call
DragerService.
Red
Inop.
LED
is
lit.
Continous
tone
commences,
also
Err
display
in
a
module
(except
air
temperature
control
module).
Fault
in
specific
module:
Skin
temperature
control
module/
humidity
control
module.
Switch
off
specific
module.
All
other
functions
will
remain
operational.
Call
DragerService.
Red
N
LED
is
lit.
Continuous
tone
commences.
No
mains
supply.
Fault
in
incubator.
Check
that
mains
plug
is
connected
to
power
supply.
Check
that
power
is
being
supplied.
Inform
in-house
technical
staff.
Call
DragerService.
38

Other manuals for Dräger 8000 NC

Related product manuals