Fault,
Cause,
Remedy
Air
temperature
control
module
Air
temperature
control
module
°C
LUFT/AIR
I
°C
>37
°C
V
±1
-
5
°
C
•&§
Sensor
M
Fault
Cause
Remedy
Red
±1.5
°C
LED
flashes.
Actual
value
display
flashes,
intermittent
tone
commences.
Air
temperature
in
incubator
is
deviating
from
set
value
by
more
than
±1.5
°C.
Below
set
value:
close
openings
in
canopy
and
wait
for
incubator
to
warm
up.
Above
set
value:
remove
additional
heat
sources
(lamps,
radiators,
sunlight).
Wait
until
incubator
has
cooled
down.
Red
TLED
flashes.
Actual
value
display
flashes,
intermittent
tone
commences.
Air
temperature
above
38
°C
(Range
of
values
set
up
to
37
°C)
Air
temperature
above
40
°C
(Range
of
values
set
up
to
39
°C)
Remove
additional
heat
sources
(lamps,
radiators,
sunlight).
Wait
for
incubator
to
cool
down.
Press
|
e
'
key
when
temperature
has
dropped
below
the
alarm
limit
for
temperature.
Red
S
LED
flashes.
Actual
value
display
flashes.
Continuous
tone
commences.
Fan
not
fitted.
Fan
not
rotating
or
rotating
too
slowly.
Fit
fan.
Check
that
fan
is
connected
to
drive
shaft
properly.
Call
DragerService.
Red
Sensor
LED
flashes.
Three
dashes
flash
in
centre
of
value
display.
Continuous
tone
commences.
Faulty
air
temperature
sensor.
Incubator
temperature
below
5
°C.
Temperature
sensor
disconnected.
Call
DragerService.
Wait
for
incubator
to
warm
up
fully.
Connect
temperature
sensor.
39

Other manuals for Dräger 8000 NC

Related product manuals