Fault,
Cause,
Remedy
Skin
temperature
control
module
Skin
temperature
control
module
(optional)
Fault
Cause
Remedy
Red
+0.5
°C
LED
flashes.
Actual
value
display
flashes,
intermittent
tone
commences.
Skin
temperature
deviating
from
set
value
by
more
than
±0.5
°C.
Below
set
value:
check
that
sensor
is
properly
fitted
to
patient.
Above
set
value:
measure
core
temperature
of
patient
and
inform
doctor
responsible
for
patient
immediately.
Red
Sensor
LED
flashes.
Three
dashes
flash
in
centre
of
actual
value
display.
Continuous
tone
commences.
Skin
temperature
sensor
not
properly
connected.
Faulty
skin
temperature
sensor.
Check
connection
and
put
right,
if
necessary.
Replace
sensor.
Err
flashes
in
actual
value
display.
Red
Inop.
LED
is
lit
in
main
module.
Continuous
tone
commences.
Fault
in
skin
temperature
control
module
while
in
operation.
Switch
to
air
temperature
control.
Call
DragerService.
Err
is
lit
in
actual
value
display.
Fault
in
the
skin
temperature
control
module;
skin
temperature
control
switched
off.
Other
modules
continue
to
operate
properly.
Call
DragerService.
Reference
temperature
of
36
°C
is
outside
tolerance
of
±0.1
°C
Check
when
key
is
pressed.
Temperature
measuring
unit
not
operating
correctly.
Call
DragerService.
Three
dashes
are
lit
at
the
top
in
actual
value
display.
Temperature
measured
above
42
°C.
Cooling
required.
Inform
doctor
responsible
for
patient
immediately.
Three
dashes
are
lit
at
bottom
of
actual
value
display.
Temperature
measured
below
30
°C.
Check
that
sensor
is
properly
fitted
to
patient.
40