Description
Design
Description
Design
The
heated
and
humidified
air
flows
over
the
entire
front
face
into
the
canopy.
It
is
channelled
over
the
front
door,
along
the
roof
of
the
canopy
and
then
drawn
down
past
the
back
wall
by
means
of
an
extraction
system.
The
baby
lies
in
a
still
area
with
very
little
air
flow.
Heat
loss
is
minimal.
When
the
hand
ports
are
open,
there
is
still
an
effective
hot
air
curtain
which
minimises
cooling
inside
the
incubator.
The
mattress
is
made
of
soft
foam
plastic
encased
in
film,
which
gives
a
very
low
heat
loss.
Humidity
The
incubator
air
is
humidified
hygienically
by
the
evaporation
of
water
from
a
water
supply.
Humidity
can
be
set
manually
according
to
a
1
to
10
scale.
Humidity
control
off
=
0.
Safety
features
When
it
is
switched
on,
the
incubator
carries
out
a
self¬
test
to
check
all
memories
in
the
microprocessor
control
system
and
to
establish
that
the
various
program
segments
are
running
correctly.
Actuators,
acknowledgement
signals
and
displays
are
switched
on
and
then
off
to
check
that
they
are
functioning
correctly.
This
test
is
also
repeated
every
10
minutes
during
operation.
All
modules
installed
in
the
incubator
are
tested.
An
Error
message
may
be
given
for
a
faulty
module,
even
if
it
is
switched
off.
The
heating
is
switched
off
if
operating
conditions
are
outside
permitted
limits.
An
additional
cooling
fan
provides
effective
cooling
as
soon
as
the
actual
value
of
the
air
temperature
exceeds
the
set
value.
49