Technical data
Technical data
Ambient conditions
Applies to:
Evita Infinity V500
Babylog VN500
Evita V300
During operation
Altitude Up to 3000 m (9842 ft)
Set values
Applies to:
Evita Infinity V500
Babylog VN500
Evita V300
The desired parameters can be set without loss of
accuracy using the therapy controls. The control-
led parameters – pressure, flow, volume, and O
2
concentration – can only be applied with the accu-
racy of the associated measured values.
The accuracies indicated apply only under the fol-
lowing conditions:
The device is ready for operation, see chapter
"Getting started."
Any accessories being used are approved for
the device, see the list of accessories.
The type of humidification is selected correctly
in the Start/Standby > Br. circuit/ Humidifier
dialog window.
The tolerances do not include the measurement
uncertainty of external test equipment. This infor-
mation is available on request.
O
2 concentration
T0 ... 90 Test conditions as per ISO 80601-2-12:2011,
Sec. 201.12.1.104
Adults
(Evita Infinity V500, Evita V300)
<35 s