For your safety and that of your patients
4
Product-specific safety information
Applies to:
– Evita Infinity V500
– Babylog VN500
–Evita V300
WARNING
Risk of malfunction
Unauthorized modifications to the medical de-
vice lead to malfunctions.
This medical device must not be modified un-
less authorized by the manufacturer.
WARNING
Risk of fire
The use of unapproved O
2 pressure reducers
can lead to excess pressure which can cause
a fire.
When supplying the ventilator with oxygen
from a compressed gas cylinder, only use
pressure reducers compliant with ISO 10524.
Slowly open the pressure reducer manually.
Do not use tools.
WARNING
Risk of unnoticed change in inspiratory O2
concentration
If an additional flow is delivered by an external
flow source, the actual O
2 concentration deliv-
ered may deviate from the displayed values.
Use additional monitoring, e.g., external SpO
2
monitoring, if necessary.
WARNING
Risk of patient injury
If leakages are present, e.g., with non-invasive
ventilation, the actual tidal volume may devi-
ate from the measured values for VTe and VTi.
Activate leakage compensation and monitor
the measured value for VT. Minimize or rem-
edy all leakages.
CAUTION
Risk of patient injury
An additional flow delivered by an external flow
source can affect the measured values for airway
pressure and flow.
CAUTION
Risk of malfunction
The touch screen has a sensitive surface. Dam-
age to the surface may cause the touch-sensitive
controls not to work properly.
Do not operate the screen with sharp objects.