37
Technical data
Volume monitoring
Applies to:
– Evita Infinity V500
– Evita V300
The expiratory minute volume MVe is monitored
within the set alarm limits.
The inspiratory tidal volume VTi or, when leakage
compensation is switched on, the leakage-compen-
sated tidal volume VT is monitored within the set
alarm limits.
Because the device ensures the minimum inspira-
tory tidal volume when volume-controlled ventila-
tion modes or pressure-controlled ventilation
modes with Volume Guarantee are selected, it is
not possible to set the lower alarm limit for VTi or
VT manually.
Applies to:
– Babylog VN500
The expiratory minute volume MVe is monitored in
the patient category Ped. pat. and in the case of in-
vasive ventilation in the patient category Neo.
within the set alarm limits.
The minimum tidal volume is monitored only when
Volume Guarantee is activated. To accomplish this,
with leakage compensation switched off the value
VTi is monitored in the patient category Ped. pat.,
and the value VTe is monitored in the patient cate-
gory Neo.. When leakage compensation is acti-
vated, VT is generally used and the automatically
set alarm limit VT low is monitored, whereby the
limit VT low corresponds to 90 % of the selected
VT.
Electromagnetic compatibility
Applies to:
– Workstation Critical Care and Neonatal Care
The references for the tables have been updated:
– Electromagnetic environment:
Information regarding electromagnetic emis-
sions (IEC 60601-1-2, table 1)
– Electromagnetic immunity:
Information regarding electromagnetic immu-
nity (IEC 60601-1-2, tables 2, 3, and 4)
– Recommended separation distances to porta-
ble and mobile RF telecommunication devices:
Information regarding separation distances
(IEC 60601-1-2, tables 5 and 6)