INFINITY CENTRAL STATION & M300 QUICK REFERENCE GUIDE
| 18
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
TO FIND A MISPLACED M300
Activating this feature causes the M300 to sound a series of five
alternating tones (assuming M300 is on, charged, within wireless
coverage area & not admitted to the ICS Main screen).
1. Select the Telemetry defaults… button on the Main menu bar.
2. Select the Find device tab.
3. Select the number corresponding to the misplaced device.
4. Select the Find button to emit tones.
TO CHANGE BED ASSIGNMENTS AT THE ICS*
1. From Main screen select System setup button.
2. Clinical password may be needed.
3. Select Layouts tab
4. Locate the Assigned Beds slot to be changed,
5. Select Remove & select Apply if appropriate.
6. Select Assign… in appropriate slot.
7. Find the appropriate device and highlight (orange box).
8. Select Assign.
9. Select Apply.
Note: Always use caution when removing and assigning beds to
prevent accidental loss of patient monitoring.
TO CHANGE THE VIEWPORT OF A M300 PATIENT*
1. Select the Bedview of the new viewport.
2. Select the Admit/Standby… button.
3. Select the Telemetry device ID of the M300 to be moved.
4. Select Apply.
5. Select NO when the pop-up question appears, the original
viewport will be discharged.
Caution: Selecting Yes will discharge the patient.
Note: Only M300 viewport to M300 viewport can be changed (can-