INFINITY COCKPIT QUICK REFERENCE GUIDE |26
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
TO ZERO CO2
Only zero a Dräger sensor when the measurement value is suspect,
or when you are prompted to re-zero on the monitor screen blue
header bar.
1. Remove sensor from airway adapter before attempting to
zero.
o Do not breathe on sensor during calibration.
o Make sure adapter is free from secretions and moisture.
2. Select CO2 parameter box.
3. Select Zero.
o ”Zero failed” or “Zero accepted” will be displayed in blue
header bar.
TO SET RRc APNEA TIME
1. Select CO2 parameter box.
2. Select RRc Apnea time (secs).
3. Dial rotary knob to select desired setting.
4. Click rotary knob to confirm.
5. Select Apnea Archive, to select Store, Str/Rec, Record, or Off
6. Dial rotary knob to select desired setting.
7. Click rotary knob to confirm.