INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE |14
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
Alarm Pause
The monitor issues alarms for parameters that are set to ON in the
Alarm Limits table. It is not a prerequisite for the parameter to be
on display or connected
for a parameter to alarm. Turning a param-
eter’s alarm function Off prevents the parameter from triggering an
alarm.
TO PAUSE AUDIO ALARMS FOR 2 MINUTES
1. Press Alarm Silence fixed key to pause an alarm for 2 minutes.
2. Parameter box becomes solid color for grade of the alarm.
3. A banner appears at the top of the screen with the message:
Audio Paused and the countdown timer.
4. Press Alarm Silence a second time while 2 minute timer is run-
ning: timer resets to 2 minutes.
5. The alarm tone resumes if a new alarm occurs during an alarm
silence period, or if any alarm condition persists past the 2 mi-
nute silence period.
Note: When backwards compatibility is enabled Alarm Silence is for 1 mi-
nute.
ALL ALARMS OFF
1. Press All Alarms OFF key to suspend visual and audible alarms
for a default set period of time (off, 1-5 min).
2. A banner appears at the top of the screen with the message:
All Alarms Paused and the countdown timer.
3. Alarms remain suspended until you press the All Alarms OFF
fixed key again or the time-out period expires.
WARNING: Never leave a patient unattended when alarms have
been permanently turned off. Always enable alarms again as soon
as possible.