INFINITY DELTA / KAPPA QUICK REFERENCE GUIDE |24
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
TO ADJUST SpO2 SETTINGS
1. Use Rotary knob to highlight SpO2 parameter box, click.
2. Use Rotary knob to highlight Waveform Size, adjust, and click.
3. Use Rotary knob to highlight Pulse Tone Source, adjust, click.
4. Use Rotary knob to highlight Pulse Tone Volume, adjust, click.
Advanced Parameters
12 Lead ECG Monitoring
Appropriate software must be installed at the Infinity Central Sta-
tion for a diagnostic quality ECG report.
1. Add additional precordial leads.
2. Connect the MultiMed-12 Pod to the monitor Aux port.
3. Connect the existing ECG lead set to the MultiMed-12 Pod.
4. Disconnect the duplicate ECG cable from the monitor.
5. Use Rotary knob to highlight ECG Report Setup, click.
6. Enter all appropriate data, click.
7. Use Rotary knob to highlight HR parameter box, click,
8. Use Rotary knob to highlight Show All Leads, click.
9. View all leads to ensure electrodes are in good contact and pa-
tient movement is not interfering with waveform interpreta-
tion.
10. Use Rotary knob to highlight REST ECG, click, a diagnostic
quality ECG Report will print to laser printer.
Invasive Pressures
Connect Hemopod or Y-cable to monitor Aux port. Connect Inter-
mediate cable if needed and transducer.
TO ZERO TRANSDUCERS FOR HEMOPOD
1. Turn stopcocks off to patient and open to air.
2. Press “0” key on the Hemopod.
3. All open transducers will zero at the same time.