INFINITY M540 QUICK REFERENCE GUIDE |12
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
Patient Category, Admit and Discharge
Confirm Patient Category for Adult, Pediatric, or Neonatal Mode.
If Category incorrect:
1. Press the Menu function key.
2. Touch the Patient setup tab.
3. Touch Patient category and Select the appropriate category
(Adult, Pediatric, & Neonate). Each patient category has spe-
cific alarm defaults associated with it.
4. The message “Changing category will change alarm set- tings
and algorithmic processing” appears.
5. Press OK. The Patient setup dialog window closes and label
changes to indicate new patient category.
Enter Patient’s Demographic Information for Admission
1. Press the Menu function key
2. Touch the Patient setup tab.
3. Select Name, ID, and Admit date and use onscreen keyboard
to enter data.
4. Press Confirm.
Note: When connected to a Central Station data entry is easily achieved
with keyboard and mouse. Height and weight may be entered at the Infinity
Central Station.
TO DISCHARGE A PATIENT
1. Press Discharge function key.
2. Press Discharge to confirm or Cancel to resume monitoring of
patient.
3. Main Screen will show “Discharged, Touch screen to resume
monitoring”.
Note: All patient data is deleted upon discharge and default limit settings
are restored.