INFINITY M540 QUICK REFERENCE GUIDE |19
THIS GUIDE IS INTENDED
FOR REFERENCE
ONLY. REFER
TO
INSTRUCTIONS FOR USE
MANUAL.ADDITIONAL INFORMATION
TO SUSSPEND INTERVAL MODE
1. Press the NIBP key longer than two seconds, to set the
interval to Off.
2. To resume interval times > reset desired interval frequency in
the Settings tab and initiate by pressing NIBP key .
SpO2 Monitoring
The accuracy of SpO2 monitoring depends largely on the strength
and quality of the SpO2 signal.
TO CHANGE PULSE TONE VOLUME
1. Touch SpO2 parameter box > touch Settings.
2. Touch Tone Volume and select desired % volume.
TO CHANGE PULSE TONE SOURCE
1. Touch SpO2 parameter box > touch Settings.
2. Touch Tone Source and select ECG or SpO2.
Note: Refer to M540 Instructions for Use manual for further device specific
information for Masimo or Nellcor.
RRi - Impedance Respiration
TO RELEARN RESP PROCESSING
1. Touch RRi parameter box > touch Settings.
2. Touch Relearn.
3. RRi waveform is re-analyzed and saved as a reference for
counting the Resp waveform.
TO CHANGE RESP PROCESSING MODE
1. Touch RRi parameter box > touch Settings.
2. Touch Mode and change to Auto or Manual.
o Auto – Optimal breath-detection threshold calculated at the
beginning of RESP monitoring, or with RESP Relearn. In-
tended for patients with regular breathing patterns. Re-