Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 133
Alarms
RRi lead off Alarm settings
–High
–Medium
– Low (default)
– One-shot
– (off)
Assigns an alarm priority to the RRI lead-off alarm
or deactivates it.
– High: The event is treated as a high-priority
alarm.
– Medium: The event is treated as a medium-
priority alarm.
– Low: The event is treated as a persistent low
alarm.
– One-shot: The event is treated as a low priority,
single notification. The message RRi lead off
appears briefly in the header bar until the user
acknowledges the condition or the condition
disappears.
– : No optical or acoustic alarm signals are
triggered.
Archive settings
–Off (default)
–Store
– Record
–Str/Rec
These archive settings cannot be configured for this
alarm message. The archive setting follows the
general archive status for the parameter. For details
on how to change the archive settings of a
parameter, refer to page 127.
PiCCO: ART catheter
unplugged
Alarm settings
Alarm column
–High (default)
– (off)
Assigns an alarm priority to the PiCCO: ART
catheter unplugged alarm or deactivates it. The
selected alarm priority affects how the alarm event
is reported optically and acoustically (see page 112
and page 114.
– High: The event is treated as a high-priority
alarm.
– : no optical or acoustic alarm signals are
triggered.
Archive settings
–Off (default)
–Store
– Record
–Str/Rec
This archive setting cannot be configured for this
alarm message. The archive setting follows the
general archive status for mean arterial blood
pressure parameter. For details on how to change
the archive settings of a parameter, refer to
page 127.
Message Available settings Description