Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 543
Troubleshooting
Mainstream CO2
Priority Message Cause Remedy
None CO
2 calibration check
failed
The mainstream sensor
calibration procedure failed.
Contact DrägerService
®
.
None CO
2 calibration check in
progress
The mainstream sensor
calibration procedure is in
progress.
Informational message – no
action required.
None CO
2 calibration check
successful
The mainstream sensor
calibration procedure was
successful.
Informational message – no
action required.
None CO
2 calibration failed The mainstream sensor
calibration procedure was
unsuccessful.
Try again or contact
DrägerService
®
.
None CO
2 calibration in
progress
The mainstream sensor
calibration procedure is in
progress.
Informational message – no
action required.
None CO
2 calibration
successful
The mainstream sensor
calibration procedure was
successful.
Informational message – no
action required.
! %0 check airway
adapter
1)
The mainstream sensor
is not properly seated on
the adapter
There are secretions in
the adapter.
There is sensor zero drift
Make sure the mainstream
sensor is attached properly
to the adapter.
If message persists, clean
or replace the airway
adapter.
If message persists though
the airway adapter is clean,
zero the sensor.
! %0 H/W failure
1)
CO2 sensor hardware
failure.
Contact DrägerService
®
.