Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 137
Alarms
Configuring ST alarm settings
Some of the ST alarm settings described are also
available on the ST alarms page (see page 233). In
the following steps, the letters in parentheses refer
to the diagram of the Limits > ST page (see
page 135). Alarm ranges and defaults are listed
starting on page 149.
To change ST alarm settings
1 Select the Alarms... button on the main menu
bar.
2 Select the Limits tab (if not already selected).
3 Select the ST tab (H) on the right side of the
page.
4 Select the setup button in the Alarm column (C)
to activate or deactivate alarm monitoring. A
crossed-out triangle appears in the parameter
field when alarm monitoring is deactivated.
5 Select the corresponding setup button in the
Lower column (D) to adjust the lower alarm
limits.
6 Select the corresponding setup button in the
Upper column (F) to adjust the upper alarm
limits.
7 Use one of the following settings in the Archive
column (G) to determine what happens in
response to an alarm:
Off – no event is stored and no recording is
generated.
Store – stores the event for later review
(see page 140).
Record – generates a timed recording
Str/Recgenerates a timed recording and
stores the event.
8 Use the Auto set all (I) button to adjust the
alarm limits for all ST parameters (see
page 139).
9 Use the Event duration [s] button (J) to select
a time an upper ST alarm limit has to be in
violation before an alarm is triggered (see
page 111).