SpO2 and Pulse CO-Ox monitoring with Masimo SET MCable
260 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Masimo SET MCable parameter field
Parameter fields report parameter values and
indicate the alarm status of parameters. Parameter
fields can also report technical conditions such as
disconnected sensors and so on. For detailed
information regarding the content of parameter
fields for each parameter, see the chapter
"Troubleshooting" on page 509.
SpO2 parameter field (Masimo SET MCable)
The SpO2 parameter field contains the following
elements:
A SpO2 label
B Units of measure – can be
activated/deactivated
C Sensitivity mode indicator (see page 264)
D PLS (pulse) label
E PLS value
F Perfusion index label
G Perfusion index value
H Alarm off symbol when alarms are deactivated.
When alarms are activated, the alarm limits are
displayed instead.
I Message area for SpO2 messages (see
page 526)
J SpO2 saturation value
K SpO2 blip that pulsates with each detected
pulse (only when the selected pulse tone
source is SpO2 – see page 264).
Pulse CO-Ox parameter field (Masimo rainbow
SET MCable)
The Pulse CO-Ox parameter field appears in
addition to the regular SpO2 parameter field when
a Masimo rainbow SET MCable is connected that
supports parameters in addition to the standard
parameter set (SpO2, PLS, PI). The parameter
content of the parameter field is configurable (see
page 266).
The display of Pulse CO-Ox parameters
(SpHb/SpHbv, SpOC, SpMet, PVI, SpCO) is
affected by the following conditions:
Blanks appear instead of parameter values if a
sensor is connected but the parameter is not
activated on the MCable.
Asterisks (***) replace the parameter values
under the following circumstances:
A parameter is activated but an
incompatible sensor is connected
A parameter is activated but no sensor is
connected
A technical failure exists (for example, an
unplugged sensor)
NOTE
The following diagram shows a typical parameter
field layout. This layout may change when
additional parameters are put on display. For
more information, see "Parameter fields" on
page 62.
038
A
B
D
E
F
G
H
J
K
C
I
NOTE
The alarm limits for SpO2 are always visible at the
Cockpit, the M540, the ICS, and on remote
devices even if the setting Show alarm limits is
deactivated (see page 460). For PLS the limits
area in the parameter field appears blank.
If SpO2 and/or PLS alarms are deactivated, the
usual symbol appears next to the parameter
label.