Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 515
Troubleshooting
!!! HR, ASY, VF off This message appears in the alarm
message field under the following
circumstances:
– Heart rate alarms are
deactivated,
– ASY/VF alarms feature is
set to Follow HR alarm (see
page 460),
– Arrhythmia monitoring is
deactivated.
The same message also appears
under the following circumstances:
– Heart rate alarms are
deactivated,
– ASY/VF alarms feature is
set to Always on (see
page 460),
– Arrhythmia monitoring is
deactivated,
– The selected HR source is
activated and is either SpO2
or ART.
Activate the functions to
remove the message.
None Pacer off
Pacer fusion
These messages appear above the
ECG waveform when the
corresponding function is activated
or deactivated (see page 214)
Deactivate the function to
remove the message.
None Privacy
Touch Screen to
resume monitoring
This message appears in the center
of the Cockpit screen when privacy
mode has been activated
(see page 88). All patient data are
removed from the screen and are
only visible at the ICS. This function
is not available unless the patient is
also admitted at the ICS.
Take the patient out of
standby to view all the data
at the Cockpit.
None Standby
Touch Screen to
resume monitoring
This message appears in the center
of the Cockpit screen when the
Cockpit has been placed in standby
mode.
Touch the screen to resume
monitoring.
Priority Message Cause Remedy