Troubleshooting
520 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
ST
Priority Message Parameter field Problem Solution
! Cannot analyze
ST
The parameter
value is replaced
by ***
The algorithm
cannot determine
ST values due to
artifact, the
absence of normal
beats, or invalid
leads.
Perform a relearn
(see page 228).
Check electrodes; re-
apply if necessary.
Make sure the
patient’s skin is
properly prepared.
Isolate the patient
from auxiliary
equipment if
possible.
Inspect and replace
faulty cable(s) and
wire(s).
Reapply gel on
reusable electrodes
and reapply them or
replace new
disposable
electrodes.
Reapply the
electrode(s). Make
sure the patient’s
skin is properly
prepared.
If a lead or electrode
cannot be replaced,
select another ST
lead for processing.
!! ST > (alarm limit)
ST < (alarm limit)
Parameter value The parameter
value is
above/below the
set upper/lower
alarm limits.
Check the patient
and treat if
necessary.
Change the alarm
limits.