Troubleshooting
524 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
%0 coincidence
1
Parameter value The heart rate and
respiratory rate fall
within 20% of each
other.
Check the
patient and
treat if
necessary.
Check and
change the
electrode
placement if
you receive a
coincidence
message until
you obtain a
clear respira-
tion signal.
Change the de-
tection thresh-
old in manual
mode or initiate
a relearning in
auto mode.
RRi relearning LEARN Relearn is in progress Informational mes-
sage -- no action
required.
1)
%0 is a placeholder for the parameter label RRi.
Priority Message Parameter field Problem Solution