Troubleshooting
540 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
! %0 H/W failure
1)
The parameter
value is replaced by
***
IBP hardware
failure.
Check hardware and
replace if necessary.
Contact
DrägerService
®
.
!! NIBP failure The parameter
value is replaced by
***
NIBP hardware
failure.
Check hardware and
replace if necessary.
Contact
DrägerService
®
.
None %0 did not zero
1)
Parameter value Transducer
zeroing failed
because of:
excessive
signal noise
a non-static
waveform
Keep all tubing
motionless, then
rezero.
Change the
transducer.
Check stopcock,
then rezero.
!! %0 static
pressure
1)
Parameter value Static pressure
detected on a
pulsatile signal,
due to:
–a
physiological
condition
such as an
asystole
a transducer
that is closed
to the patient
a catheter tip
that is
lodged
against a
vessel wall
a clot on the
catheter tip
Check the patient
and treat if
necessary.
Open the system to
the patient by turning
the stopcock.
Follow hospital
procedures for
dislodging catheters.
Follow hospital
procedures for
clotted catheters.
1)
%0 is a placeholder for the respective IBP label (including CPP).
Priority Message Parameter field Problem Solution