Troubleshooting
540 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
! %0 H/W failure
1)
The parameter
value is replaced by
***
IBP hardware
failure.
– Check hardware and
replace if necessary.
– Contact
DrägerService
®
.
!! NIBP failure The parameter
value is replaced by
***
NIBP hardware
failure.
– Check hardware and
replace if necessary.
– Contact
DrägerService
®
.
None %0 did not zero
1)
Parameter value Transducer
zeroing failed
because of:
– excessive
signal noise
– a non-static
waveform
– Keep all tubing
motionless, then
rezero.
– Change the
transducer.
– Check stopcock,
then rezero.
!! %0 static
pressure
1)
Parameter value Static pressure
detected on a
pulsatile signal,
due to:
–a
physiological
condition
such as an
asystole
– a transducer
that is closed
to the patient
– a catheter tip
that is
lodged
against a
vessel wall
– a clot on the
catheter tip
– Check the patient
and treat if
necessary.
– Open the system to
the patient by turning
the stopcock.
– Follow hospital
procedures for
dislodging catheters.
– Follow hospital
procedures for
clotted catheters.
1)
%0 is a placeholder for the respective IBP label (including CPP).
Priority Message Parameter field Problem Solution