Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 553
Troubleshooting
None C.O. injectate set to
20°C!
Parameter value No thermistor was
connected. The
M540 assumes a
temperature of
20 °C (68 °F).
Attach a thermistor.
None %0 No Temperature
Change
1)
Parameter value No change in
blood temperature
during the C.O.
measurement.
Repeat the
measurement.
Use a larger
injectate volume.
Repeat the
measurement. If
problem persists,
replace the
catheter.
Use a cooler
injectate.
None %0 Poor Baseline
1)
Parameter value Poor blood
temperature
baseline during
C.O.
measurement.
Follow hospital
procedures.
Repeat the
measurement.
Replace the faulty
components, if the
message persists.
None %0 Use Cooler
Injectate
1)
Parameter value The difference
between the
temperature of
the blood and
the injectate is
less than 5 °C
(41 °F).
The injectate
temperature is
greater than
25 °C (77 °F).
Use a colder injectate.
1)
%0 is a placeholder for the parameter label C.O.
Priority Message Parameter field
Problem Solution