Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n 565
Troubleshooting
Medium Scio is not connected Scio is disconnected or turned
off.
Connect the Scio module
or turn it on.
Low Scio unavailable for
neonates
– Scio is plugged in while
IACS is already in neonate
mode.
– IACS is switched to
neonate mode while Scio
is already plugged in.
– Connect an alternate
CO
2 monitor (e.g.,
Mainstream or
Microstream) if CO2
monitoring is desired
in neonate mode.
– Switch IACS out of
neonate mode in
order to continue
Scio monitoring.
Low Scio warming up:
Accur. low
Accuracy is not guaranteed
while the Scio is warming up.
Wait for the Scio module
to warm up.
Low Second agent
detected
NOTE
This alarm occurs only
on gas analyzers with
automatic agent
identification.
A second anesthetic agent
has been detected.
NOTE
This alarm could be an
expected clinical behavior if
the clinician regularly uses
two agents as part of the
process.
– Wait for the transition
phase to end after
changing anesthetic
agents.
– Flush the system if
necessary.
– Check fresh-gas
settings.
Medium Third agent detected
NOTE
This alarm occurs only
on gas analyzers with
automatic
identification.
A mixture of three or more
anesthetic agents or other
gases has been detected,
possibly as a result of:
– A change of the anesthetic
agent during monitoring
– Electromagnetic
interference
– The use of inhalants or
sprays (e.g., albuterol)
– Wait for the transition
phase to end after
changing anesthetic
agents.
– Flush the system if
necessary.
– Check fresh-gas
settings.
– Check for
electromagnetic
radiation in the
vicinity.
Low Water trap is full – Water trap is full.
– Sample line is occluded.
– Check water trap.
– Check sample line,
water trap, and
patient-side filter.
Alarm Priority Alarm Cause Remedy